Regular bump testing is a critical component of maintaining safety when working with gas detection instruments, especially in environments where hazardous gases may be present. A bump test provides a quick and essential check to ensure your gas detector is operational and capable of alerting personnel to potential dangers. Here's a comprehensive guide to understanding and performing a proper bump test on a ground gas detection device.
1. What Is a Bump Test and Why It Matters
A bump test is a functional test that briefly exposes a gas detector to a known concentration of gas to verify that its sensors respond appropriately and that all alarm systems—visual, audible, and vibratory—are activated. Unlike calibration, which adjusts sensor accuracy, a bump test simply confirms the device is operational. It's a frontline safety measure designed to catch sensor or alarm malfunctions before you enter a potentially dangerous environment.
2. Essential Equipment You’ll Need
Before beginning, gather the necessary components:
-
Gas detection instrument (specific to your monitoring requirements, such as methane, hydrogen sulfide, or oxygen)
-
Bump test gas cylinder containing a certified, known concentration of target gases (ensure it is within its expiration date)
-
Regulator and tubing, compatible with your gas cylinder
-
Bump test cap or adapter, usually supplied by the instrument manufacturer
Safety Note: If your bump gas contains flammable or toxic gases (e.g., methane or carbon monoxide), always perform the test in a well-ventilated, controlled environment—ideally outdoors or in a designated testing area.
3. Preparing the Instrument for Testing
Turn on the gas detection device and allow it to complete its startup diagnostics and warm-up period. This typically takes a few minutes and ensures the sensors are functioning under normal operating conditions. Always follow the manufacturer’s specific warm-up instructions for your model.
4. Properly Connecting the Bump Test Gas
Carefully attach the regulator to the bump test gas cylinder. Connect the tubing to the output of the regulator and the other end to the bump test cap or directly to the instrument’s gas inlet, depending on your model. Ensure all fittings are tight and leak-free. Leaks can result in inaccurate testing and expose personnel to hazardous gas.
5. Applying the Bump Gas to the Sensor
Secure the bump test cap or adapter over the gas sensor or inlet port. Slowly open the regulator valve to allow gas to flow. This should typically take 10–30 seconds, depending on the sensor response time and flow rate (usually between 0.5–1.0 LPM). The gas detector should register the gas and begin displaying real-time readings.
6. Monitoring Instrument Response
Carefully observe how the device reacts:
-
Does the display show the expected gas concentration range?
-
Do all alarms activate—audible, visual (LEDs), and vibratory (if equipped)?
-
Are the sensors responding within an acceptable timeframe (as defined by the manufacturer)?
If the instrument does not respond appropriately, do not use it until the issue is identified and resolved.
7. Recordkeeping and Compliance
Document the results of the bump test in your maintenance log. Record the following:
-
Date and time of the test
-
Operator’s name
-
Instrument model and serial number
-
Bump gas concentration and expiration date
-
Instrument’s reading and whether alarms triggered
Maintaining accurate bump test records supports compliance with workplace safety standards such as OSHA, AS/NZS 60079.29.2, or EN 60079 standards, and proves due diligence in safety audits.
8. Troubleshooting: What If the Test Fails?
If the instrument fails to respond during the bump test, consider the following common causes:
-
Expired or incorrect bump test gas – Always verify the gas cylinder’s shelf life and composition.
-
Obstructed or leaking tubing – Check connections and tubing for cracks or loose fittings.
-
Sensor contamination or failure – Over time, sensors can become contaminated by dust, moisture, or corrosive gases.
In most cases, cleaning the sensor or replacing the gas cylinder resolves the issue.
9. When to Call for Professional Support
Persistent test failures may indicate a deeper hardware or sensor issue. If you've ruled out common faults and the instrument still doesn't perform, contact the manufacturer or a certified calibration service provider. Instruments operating in high-risk environments should only be used when fully functional and validated.