Temperature Process Calibrator - Simulation/Measurement - SPMK516
Temperature Process Calibrator - Simulation/Measurement - SPMK516
The SPMK516 is an essential instrument for all field and in house calibration tasks. It meets the requirements of Calibration, Maintenance and Repair of thermal instruments. It is the ideal instruments for both field configuration, fault diagnosing and calibration tasks. With efficiently Source or Measure Millivolt output with High Precision and Stable Readings in Seconds.
Simple set up, plug in and play interface. Featuring calibrations for various instruments such as Thermocouples, RTD, Temperature Controllers, Pressure Controllers. Making it ideal for metrology institutes and enterprises seeking to establish calibrations.
Main Features
◼ Measure voltage, current, resistance, frequency, pulse, switch (on and off),also can use RTD / thermocouple measuring temperature, which are mutual interference.
◼ Simulate 13 kinds of thermocouple and 11 kinds of RTD measure and output, with cold junction compensation.
◼ DC24V power supply isolated from measurement and output.
◼ AC/DC power supply: Battery power is recommended for better performance.
Collapsible content
Functions
Simulating
Millivolt voltage
Voltage
Current
Frequency
Pulse
Resistance
Simulate 11 RTDs
Simulate 13thermocouples
Measurement
Millivolt voltage
Voltage
Current
Frequency
Pulse
Resistance
11 RTDs
13 thermocouples
Other Functions
Thermal book
Step programming output
24V loop power supply
Simulating transmitter
RS232 communication
Storage
Techincal Data
Electrical Specifications—Source (Ambient temperature (20±5)°C,1 year accuracy)
Signal | Range | Accuracy | Resolution |
---|---|---|---|
Current | (0-22)mA | ±(0.02%RD+0.005%F.S) | 0.1μA |
Voltage | (0-12)V | ±(0.02%RD+0.005%F.S) | 0.1mV |
mV | (-10-100)mV | ±(0.02%RD+0.005%F.S) | 0.1μV |
Resistance | (0-400)Ω | ±(0.02%RD+0.005%F.S) | 1mΩ |
Resistance | (400-4000)Ω | ±(0.03%RD+0.01%F.S) | 10mΩ |
Frequency | (1-50000)Hz | ±(0.005%RD+0.002%F.S) | 0.1Hz |
Pulse | 0-999999 | ±1pcs | 1 |
DC24V | 24.0±0.5V | ||
Thermocouple | K E J T N R S B C D G L U | ||
RTD |
Pt1000(385), Pt500(385), Pt100(385), Pt100(3916), Pt100(3926), Pt10(385) Cu10(427), Cu50(428), Cu100(428), Ni120(672), Ni100(618) |
Electrical Specifications—Measurement (Ambient temperature (20±5)°C,1 year accuracy)
Signal | Range | Accuracy | Resolution |
---|---|---|---|
Current | (-30-30)mA | ±(0.01%RD+0.005%F.S) | 0.1μA |
Voltage | (-30-30)V | ±(0.01%RD+0.005%F.S) | 0.1mV |
mV | (-75-75)mV | ±(0.01%RD+0.005%F.S) | 0.1μV |
R-3W |
(0-400)Ω (0-4000)Ω |
±(0.02%RD+0.005%F.S) |
1mΩ 10mΩ |
R-4W |
(0-400)Ω (0-4000)Ω |
±(0.01%RD+0.005%F.S) 1mΩ |
1mΩ 10mΩ |
Frequency | (1-50000)Hz | ±(0.005%RD+0.002%F.S) | 1Hz |
Pulse | 0-999999 | ±1pcs | 1 |
Limit Switch | If the switch has detective voltage, range is (3-24)V | ||
Thermocouple | K E J T N R S B C D G L U | ||
RTD |
Pt1000(385), Pt500(385), Pt100(385), Pt100(3916), Pt100(3926), Pt10(385) Cu10(427), Cu50(428), Cu100(428), Ni120(672), Ni100(618) |
Shipping
We aim to process and get your delivery with you as quick as possible -
Shipping rates and delivery times are calculated at checkout.
Guarantee
We aim to provide you with the best quality products, if for an unfortunate product defect we have a 14 day money back return guarantee.
Please see our returns policy for more information.